Opening the laboratory controls…Reload this pageAll laboratories

LFM LAB / PUT THE FOUR CONCEPTS TO WORK

First independent experiment

Resolve a close bead pair, capture a clear image, and report a calibrated distance.

Return to the full LFM lab ↗

Operate the microscope

FOLLOW THE LIGHTExcitation in. Fluorescence out.
View options

The simple diagram shows the same light route without perspective.

Shutter closed · illustrated pathDashed guides show where light would travel

Preparing the optical assembly…

01 · Excitation source
Explore the hardware · Excitation source

Excitation source. Provides widefield excitation. The illuminated field is not raster-scanned.

Illustrative cells; not a one-to-one match to the microscope image.

Blue: excitationGreen: returning fluorescenceDashed paths: illustration only

Expanded teaching layout · nearby hardware omitted in close-ups · relay optics condensed · directions and scanning exaggerated. Mounts and glass are illustrative, not a manufacturer’s instrument.

Blue: excitation → specimenGreen: emission → camera

Shutter closed: light stops at the shutter. Faint lines mark the available route.

Arrows show the direction of light; the assembly stays still for inspection. The detector preview predicts the next exposure.

Working principle · 1. Excitation source

1. Excitation source

Provides the excitation light. Increasing intensity increases the photon rate, not the speed of light. This widefield model does not simulate a focused laser scanning the specimen.

Current controls: 20× / 0.45 · air · focus 1.50 µm · 8% excitation · 80 ms exposure · detector gain 1

Expanded cutaway; excitation and emission are slightly separated in the drawing to show their opposite directions through the same objective; packet speed, count and size are illustrative. Specimen focus motion is enlarged. The condenser beneath the slide belongs to transmitted illumination and is not used by this epi-fluorescence route.

Predicted preview · shutter closed

Close-pair detail · enlarged 32 × 32 pixel crop.
Display enlargement adds no optical resolution.

Upper pair: 20 µm reference · middle: measure its centres · bottom: 0.70 µm resolution challenge.

Only live illumination and capture add simulated dose. This preview predicts an exposure while the shutter is closed.

Rayleigh reference 0.705 µmPredicted field 130.0 µmTeaching dose 0.000Saved peak No captureSaved clipping No capture

Measure the saved capture

Capture settings stay fixed here when you adjust the microscope. A new capture replaces this image and clears its calibration.

Capture an image to unlock calibration and measurement.

Explain your choices

About this simulation

This exercise uses the LFM lab’s synthetic fluorescence image engine: Gaussian emitter profiles broaden with defocus, photon and read noise affect raw counts, and clipping occurs before display mapping. The original six-bead phantom is in one plane, with 520 nm emission. Optical estimates are idealized. Detector gain scales counts before ADC clipping; the image display itself stays fixed. The 3D instrument is an explanatory schematic; measurements use saved 512 × 512 source pixels. The dose budget is a teaching constraint, not a physical damage limit. Pass requires all six criteria. The 20 µm reference and the middle pair must each be measured centre-to-centre.